Back to Search
Start Over
Comparison of a hydroxyapatite-coated sleeve and a porous-coated sleeve with a modular revision hip stem. A prospective, randomized study.
- Source :
- Journal of Bone & Joint Surgery, American Volume; Dec2004, Vol. 86 Issue 12, p2720-2725, 6p
- Publication Year :
- 2004
-
Abstract
- <bold>Background: </bold>Bone ingrowth into a cementless prosthesis can be achieved by both porous and hydroxyapatite coatings. The purpose of this study was to compare the performance of a hydroxyapatite-coated proximal sleeve and a porous bead-coated sleeve in patients managed with a modular revision hip system.<bold>Methods: </bold>Between August 1992 and December 1996, fifty-three consecutive femoral revisions performed with an S-ROM stem in fifty-two patients were prospectively randomized at the time of surgery to either a hydroxyapatite-coated or a porous-coated sleeve. All patients were evaluated clinically and radiographically at three months, six months, and yearly for a minimum of two years (average, four years; range, two to 7.5 years). Femoral defects were classified according to the criteria of Paprosky et al. Six patients died and four patients were lost to follow-up, leaving forty-two patients (forty-three hips) as the final study group.<bold>Results: </bold>For the entire group, two femoral stems, one of which had been implanted in a hip with a Paprosky type-II femoral defect and the other in a hip with a Paprosky type-IIIB femoral defect, required a repeat revision, one for pain and the other for aseptic loosening. Radiographic evidence of bone ingrowth was observed in 96% (twenty-six) of the twenty-seven femora with type-I or II defects and in 81% (thirteen) of the sixteen femora with type-III defects. Femoral component survival, with use of revision as the end point, was 95% at four years for the entire group. The Harris hip scores were not significantly different when stratified by implant type, but were significantly different when stratified by bone loss (p < 0.05). In the femora with type-I or II defects, no difference was detected between those treated with a hydroxyapatite-coated implant and those that received a porous-coated implant with respect to bone ingrowth. However, in femora with type-III defects, the likelihood of the development of bone ingrowth was 2.6 (95% confidence interval, 1.3 to 5.17) times greater in hips that received a hydroxyapatite-coated implant (all eight developed ingrowth) than in hips that had a porous-coated implant (five of eight developed ingrowth) (p = 0.05).<bold>Conclusions: </bold>Bone fixation was achieved more often with hydroxyapatite-coated sleeves in femora with Paprosky type-III defects, but no significant difference was noted in outcomes between the two implant types when used in bone with type-I or type-II femoral defects. Overall, the S-ROM modular hip stem performed better in femora with type-I or II bone defects than in femora with type-III defects.<bold>Level Of Evidence: </bold>Prognostic study, Level II-1 (retrospective study). See Instructions to Authors for a complete description of levels of evidence. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00219355
- Volume :
- 86
- Issue :
- 12
- Database :
- Complementary Index
- Journal :
- Journal of Bone & Joint Surgery, American Volume
- Publication Type :
- Academic Journal
- Accession number :
- 106503439