Back to Search
Start Over
Homogeneous Liquid-Liquid Extraction Coupled to Ion Mobility Spectrometry for the Determination of p-Toluidine in Water Samples.
- Source :
- Bulletin of Environmental Contamination & Toxicology; Apr2015, Vol. 94 Issue 4, p474-478, 5p, 3 Charts, 3 Graphs
- Publication Year :
- 2015
-
Abstract
- In this research, homogeneous liquid-liquid extraction followed by ion mobility spectrometry (HLLE-IMS) with corona discharge ionization source has been developed for the determination of p-toluidine. The analyte was extracted by single-phase extraction in a ternary solvent system and then the extracted p-toluidine was injected into IMS for analysis. Optimization of different parameters which could influence HLLE-IMS was performed. Under optimum conditions, the dynamic linear range was obtained over 2.0-40.0 ng/mL with R = 0.9966 while relative standard deviation was below 10 %. The limits of detection and quantification were 0.6 and 2.0 ng/mL of p-toluidine, respectively. The proposed method was applied to determine p-toluidine in environmental water samples which resulted in acceptable recoveries of the analyte, ranging from 85.3 %-90 %. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 00074861
- Volume :
- 94
- Issue :
- 4
- Database :
- Complementary Index
- Journal :
- Bulletin of Environmental Contamination & Toxicology
- Publication Type :
- Academic Journal
- Accession number :
- 101423141
- Full Text :
- https://doi.org/10.1007/s00128-014-1426-z