Back to Search Start Over

Simultaneous Determination of Nitrite and Hydrogen Peroxide Using Hemoglobin Modified Pencil Lead Electrode as a Novel Biosensor: Application to the Analysis of Mother's Milk.

Authors :
Majidi, Mir Reza
Pournaghi‐Azar, Mohammad Hossein
Saadatirad, Afsaneh
Alipour, Esmaeel
Source :
Journal of the Chinese Chemical Society; Jan2015, Vol. 62 Issue 1, p83-89, 7p
Publication Year :
2015

Abstract

In the present paper simultaneous determination of nitrite and hydrogen peroxide using hemoglobin modified pencil lead electrode (Hb/PLE) prepared by a simple and rapid electroless method was described. In the first part of the work the reduction of NO<subscript>2</subscript> and H<subscript>2</subscript>O<subscript>2</subscript> at the modified electrode was investigated by cyclic voltammetry. Then under optimal conditions using differential pulse voltammetry, the biosensor could be used for the determination of H<subscript>2</subscript>O<subscript>2</subscript> at concentration ranging from 5 to 240 M and NO<subscript>2</subscript> at concentration ranging from 10 to 240 M. The detection limits were 3×10<superscript>−6</superscript> and 5×10<superscript>−6</superscript> M, for NO<subscript>2</subscript> and H<subscript>2</subscript>O<subscript>2</subscript> respectively. Differential pulse voltammetry also used for the simultaneous determination of NO<subscript>2</subscript> and H<subscript>2</subscript>O<subscript>2</subscript>. This modified electrode successfully used for the determination of NO<subscript>2</subscript> and H<subscript>2</subscript>O<subscript>2</subscript> in tap water and mother's milk samples. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00094536
Volume :
62
Issue :
1
Database :
Complementary Index
Journal :
Journal of the Chinese Chemical Society
Publication Type :
Academic Journal
Accession number :
100488260
Full Text :
https://doi.org/10.1002/jccs.201400155