Quantitative assay for epinephrine in dental anesthetic solutions by capillary electrophoresis.

Authors :
Britz-Mckibbin P
Kranack AR
Paprica A
Chen DD
Source :
The Analyst [Analyst] 1998 Jul; Vol. 123 (7), pp. 1461-3.
Publication Year :
1998

Abstract

A simple and robust method for the separation and quantification of epinephrine in dental anesthetic solutions was developed. The method allows the direct injection of high salt solutions without sample pre-treatment. Large sample plugs (5.7% of the total capillary length) are used for epinephrine determination by selective analyte focusing in capillary electrophoresis. The concentration detection limit for epinephrine is about 5.0 x 10(-7) M (90 ng ml-1) with a commercial UV detector. The separation protocol was validated in terms of its precision, linearity, accuracy and specificity.

Details

Language :
English
ISSN :
0003-2654
Volume :
123
Issue :
7
Database :
MEDLINE
Journal :
The Analyst
Publication Type :
Academic Journal
Accession number :
9830160
Full Text :
https://doi.org/10.1039/a800772a