Validation of the Doser, a new device for monitoring metered-dose inhaler use.
- Source :
-
The Journal of allergy and clinical immunology [J Allergy Clin Immunol] 1998 Sep; Vol. 102 (3), pp. 409-13. - Publication Year :
- 1998
-
Abstract
- Background: Electronic monitoring of medication use has proved valuable in both clinical and research settings. The Doser, a new and inexpensive commercially available device for monitoring metered-dose inhaler (MDI) use, displays 3 measures of daily use of an attached MDI: (1) the daily total of actuations, (2) the number of doses remaining in the MDI, and (3) the number of actuations on each of the preceding 30 days for later recall.<br />Objective: We sought to validate the accuracy of the Doser with several commonly prescribed MDIs.<br />Methods: In the laboratory, clinic personnel actuated an MDI with an attached Doser several times in succession on 3 consecutive days and recorded each of the 3 measures of MDI use (study 1). In study 2 clinic personnel carried an MDI and attached Doser with them for 4 weeks, actuating the MDI according to a prescribed protocol each morning and evening and again recording each of the 3 measures of daily use. In addition, during 2 weeks of study 2, a thermistor-based Nebulizer Chronolog was attached to the MDI to electronically record the date and time of each actuation. In study 3 clinic patients had both a Doser and Nebulizer Chronolog attached to their routinely used inhalers for 2 weeks and a Doser alone during a separate 2-week period.<br />Results: In study 1 agreement was 99% to 100% among the 3 Doser measures, and each measure agreed with actual use by self-report 97% of the time. In study 2 agreement among the 3 Doser measures of use ranged from 98% to 99%. Agreement between each of the 3 Doser measures and the Nebulizer Chronolog ranged from 90% to 93%. Agreement between each of the 3 Doser measures and actual use ranged from 96% to 97%, and the Nebulizer Chronolog agreed with actual use 93% of the time. In study 3 Doser and Nebulizer Chronolog agreement with patient self-report were 85% and 80%, respectively. Agreement between the Doser and Nebulizer Chronolog was 76%. Several failures of the thermistor-based Nebulizer Chronolog occurred, and occasional mechanical problems occurred with the Doser, primarily on particular types of MDI canisters.<br />Conclusion: The Doser provides an accurate measure of MDI use with most commonly prescribed medications and may be useful for monitoring MDI use by investigators, clinicians, and patients.
Details
- Language :
- English
- ISSN :
- 0091-6749
- Volume :
- 102
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- The Journal of allergy and clinical immunology
- Publication Type :
- Academic Journal
- Accession number :
- 9768581
- Full Text :
- https://doi.org/10.1016/s0091-6749(98)70128-9