Back to Search Start Over

[Pitfalls when using a contrast media injector in MRI].

Authors :
Heverhagen JT
Hoppe M
Ishaque N
Froelich JJ
Klose KJ
Source :
RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin [Rofo] 1998 Aug; Vol. 169 (2), pp. 198-200.
Publication Year :
1998

Abstract

Purpose: Using a power injector to applicate Gd-DTPA we found a contrast enhancement of the pyelo-caliceal system even in the native studies, and hence we analysed pitfalls when using power injectors in MRI.<br />Material and Methods: We used a power injector Spectris (MedRad, Maastricht, Netherlands). In vitro artifacts were achieved by the mixture of contrast media and saline solution. We substituted contrast media by red water, NaCl by clear water.<br />Results: Using power injectors in MRI, some pitfalls must be avoided, which can render investigations useless, especially dynamic contrast-enhanced investigations.<br />Conclusion: In our study we showed an easy way to overcome some pitfalls and use a power injector in MRI in a diagnostically helpful way. The simple use of valves inhibits the mixture of contrast media and saline solution.

Details

Language :
German
ISSN :
1438-9029
Volume :
169
Issue :
2
Database :
MEDLINE
Journal :
RoFo : Fortschritte auf dem Gebiete der Rontgenstrahlen und der Nuklearmedizin
Publication Type :
Academic Journal
Accession number :
9739373
Full Text :
https://doi.org/10.1055/s-2007-1015074