Successful redeployment of an unexpanded coronary stent.

Authors :
McGinnity JG
Glazier JJ
Spears JR
Rogers C
Turi ZG
Source :
Catheterization and cardiovascular diagnosis [Cathet Cardiovasc Diagn] 1998 May; Vol. 44 (1), pp. 52-6.
Publication Year :
1998

Abstract

We report on treatment of a patient in whom failure to deploy the distal portion of a Palmaz-Schatz stent occurred but was not recognized. After an unstable course, the patient underwent repeat coronary angiography, at which time the stent was rewired and redilated. Full deployment of the stent with restoration of TIMI grade 3 flow was achieved. The putative cause of the problem, incomplete deployment of the stent because of inadvertent advancement of the stent delivery sheath, should be avoided, and needs to be recognized if it occurs. Crossing and redilating the stent is possible, although technically difficult.

Details

Language :
English
ISSN :
0098-6569
Volume :
44
Issue :
1
Database :
MEDLINE
Journal :
Catheterization and cardiovascular diagnosis
Publication Type :
Academic Journal
Accession number :
9600524
Full Text :
https://doi.org/10.1002/(sici)1097-0304(199805)44:1<52::aid-ccd13>3.0.co;2-r