Back to Search
Start Over
Timing algorithm for bolus chase MR digital subtraction angiography.
- Source :
-
Magnetic resonance in medicine [Magn Reson Med] 1998 May; Vol. 39 (5), pp. 691-6. - Publication Year :
- 1998
-
Abstract
- To acquire multiple longitudinal locations in the lower extremity after a single contrast injection, appropriate table translation and contrast injection are required. An approximate model based on constant bolus velocity was developed to describe the space-time course of a contrast bolus in the lower extremity. This model was verified in dynamic MR angiograms acquired in a group of patients using time-resolved 2D MR digital subtraction angiography (MRDSA). From this contrast bolus passage model, a timing algorithm for table translation and contrast injection was developed for bolus chase MRDSA, subsequently validated in bolus chase 2D MRDSA experiments. All targeted major peripheral arteries were well depicted in bolus chase 2D MRDSA using this timing algorithm and a single 15-ml contrast dose.
Details
- Language :
- English
- ISSN :
- 0740-3194
- Volume :
- 39
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Magnetic resonance in medicine
- Publication Type :
- Academic Journal
- Accession number :
- 9581598
- Full Text :
- https://doi.org/10.1002/mrm.1910390504