Back to Search Start Over

Timing algorithm for bolus chase MR digital subtraction angiography.

Authors :
Wang Y
Lee HM
Avakian R
Winchester PA
Khilnani NM
Trost D
Source :
Magnetic resonance in medicine [Magn Reson Med] 1998 May; Vol. 39 (5), pp. 691-6.
Publication Year :
1998

Abstract

To acquire multiple longitudinal locations in the lower extremity after a single contrast injection, appropriate table translation and contrast injection are required. An approximate model based on constant bolus velocity was developed to describe the space-time course of a contrast bolus in the lower extremity. This model was verified in dynamic MR angiograms acquired in a group of patients using time-resolved 2D MR digital subtraction angiography (MRDSA). From this contrast bolus passage model, a timing algorithm for table translation and contrast injection was developed for bolus chase MRDSA, subsequently validated in bolus chase 2D MRDSA experiments. All targeted major peripheral arteries were well depicted in bolus chase 2D MRDSA using this timing algorithm and a single 15-ml contrast dose.

Details

Language :
English
ISSN :
0740-3194
Volume :
39
Issue :
5
Database :
MEDLINE
Journal :
Magnetic resonance in medicine
Publication Type :
Academic Journal
Accession number :
9581598
Full Text :
https://doi.org/10.1002/mrm.1910390504