Multicenter evaluation of the Paragon CZE 2000 capillary zone electrophoresis system for serum protein electrophoresis and monoclonal component typing.
- Source :
-
Clinical chemistry [Clin Chem] 1998 Mar; Vol. 44 (3), pp. 599-605. - Publication Year :
- 1998
-
Abstract
- Serum protein electrophoresis and typing of monoclonal components (MCs) are routine but time-consuming and technically demanding assays. We evaluated capillary electrophoresis (Paragon CZE 2000) for automation of the two assays. CZE and cellulose acetate electrophoresis gave similar data on 794 samples. Within-run and between-run CVs were < 2% for albumin and gamma-globulins and 4-7% for alpha 1-, alpha 2-, and beta-globulins. Bilirubin, hemoglobin, triglycerides, and fibrinogen were found not to interfere. No carryover by capillaries was detected. The detection limit for MC was < 0.5 g/L. MC assessment by immunosubtraction on 403 samples identified the monoclonal type in all samples with peak concentrations > 10 g/L; only 50% of MCs that could not be quantified by densitometric scan were typed.
- Subjects :
- Bilirubin blood
Electrophoresis, Capillary methods
Electrophoresis, Cellulose Acetate methods
Fibrinogen
Hemoglobins
Humans
Immunoelectrophoresis methods
Laboratories standards
Nephelometry and Turbidimetry methods
Nephrotic Syndrome blood
Quality Control
Reference Values
Reproducibility of Results
Sensitivity and Specificity
Serum Albumin isolation & purification
Serum Globulins isolation & purification
Triglycerides blood
gamma-Globulins isolation & purification
Antibodies, Monoclonal
Blood Proteins isolation & purification
Electrophoresis, Capillary instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 0009-9147
- Volume :
- 44
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Clinical chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 9510868