High-performance liquid chromatography-tandem mass spectrometry as a reference for analysis of tacrolimus to assess two immunoassays in patients with liver and renal transplants.
- Source :
-
Therapeutic drug monitoring [Ther Drug Monit] 1997 Dec; Vol. 19 (6), pp. 694-700. - Publication Year :
- 1997
-
Abstract
- The accuracy and imprecision of three assays used for therapeutic monitoring of tacrolimus were tested using blood-containing weighed-in amounts of the drug, an enzyme-linked immunosorbent assay (ELISA), a microparticle enzyme immunoassay (MEIA I), and a high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS2) assay. Accuracy was acceptable for the HPLC-MS2 assay at all concentrations tested (< 10% deviation) and for the ELISA at 1.0 and 4.0 microg/l. Accuracy was not acceptable for the ELISA at 15.0 and 50.0 microg/l or for the MEIA I at all concentrations tested. Imprecision was acceptable for the HPLC-MS2 assay at all concentrations tested (coefficient of variation < 10%), for the ELISA at 15.0 and 50.0 microg/l, and for the MEIA I at 15.0 and 50.0 microg/l. Imprecision was not acceptable for the ELISA at 1.0 and 4.0 microg/l or for the MEIA I at 1.0 and 4.0 microg/l. This assessment with weighed-in amounts of tacrolimus verified the HPLC-MS2 assay as a reference method. The performance of the two immunoassays with HPLC-MS2 was then compared in the clinical setting using blood from patients with liver (n = 30) and renal (n = 37) transplants. In the liver transplant group (127 samples), the range of tacrolimus concentrations measured by HPLC-MS2, ELISA, and MEIA I was 1.9 to 31.8, 2.1 to 35.0, and less than 0.1 to 36.5 mg/l, respectively. In the renal transplant group (129 samples), the ranges were 1.7 to 26.1, 1.9 to 24.4, and 0.9 to 28.5 microg/l, respectively. Compared with the HPLC-MS2, the ELISA had minimal bias (0.1 to 0.2 microg/l) but unacceptable variability in values (SD > 13%). The MEIA I had unacceptable bias (1.7-1.8 microg/l) and variability (SD > 23%). These data indicated that neither the ELISA nor MEIA I is interchangeable with HPLC-MS2. Moreover, in view of the current trend to reduce the therapeutic dose of tacrolimus, quantitative results using the MEIA I would not be obtainable during therapeutic drug monitoring in some patients in whom effective therapeutic concentrations can be less than 5.0 microg/l.
Details
- Language :
- English
- ISSN :
- 0163-4356
- Volume :
- 19
- Issue :
- 6
- Database :
- MEDLINE
- Journal :
- Therapeutic drug monitoring
- Publication Type :
- Academic Journal
- Accession number :
- 9421113
- Full Text :
- https://doi.org/10.1097/00007691-199712000-00015