Back to Search Start Over

Analysis of urinary pseudouridine by micellar electrokinetic capillary chromatography.

Authors :
Evans MD
Perrett D
Lunec J
Herbert KE
Source :
Annals of clinical biochemistry [Ann Clin Biochem] 1997 Sep; Vol. 34 ( Pt 5), pp. 527-33.
Publication Year :
1997

Abstract

We describe a capillary electrophoresis procedure, using uridine as an internal standard, for the analysis of urinary pseudouridine following solid-phase extraction. This method retains the advantages of existing chromatographic techniques but has superior resolving power and is technically less demanding. The standard curves were linear and reproducible with a detection limit of 60 fmol; chromatographic analysis was complete in under 10 min. Injection variability was < 5% and multiple independent analyses of the same urine sample for pseudouridine concentration gave coefficients of variation of < 10%. The mean (SD) urinary pseudouridine level in 18 healthy subjects was 16.1 (2.1) nmol/mumol creatinine. For a limited group of subjects where samples were taken more frequently, intra-individual variation averaged 27.5% reflecting variable excretion.

Details

Language :
English
ISSN :
0004-5632
Volume :
34 ( Pt 5)
Database :
MEDLINE
Journal :
Annals of clinical biochemistry
Publication Type :
Academic Journal
Accession number :
9293307
Full Text :
https://doi.org/10.1177/000456329703400506