Analysis of terfenadine in human plasma using microbore high-performace liquid chromatography-electrospray ionisation mass spectrometry.

Authors :
Adams DA
Murray S
Clifford CP
Rendell NB
Davies DS
Taylor GW
Source :
Journal of chromatography. B, Biomedical sciences and applications [J Chromatogr B Biomed Sci Appl] 1997 Jun 06; Vol. 693 (2), pp. 345-51.
Publication Year :
1997

Abstract

An assay based on combined microbore high-performance liquid chromatography-positive ion electrospray ionisation mass spectrometry with selected ion recording has been developed for the measurement of the antihistamine drug terfenadine in human plasma. A deuterated analogue of terfenadine was synthesised for use as an internal standard and extraction of terfenadine was carried out on C18 solid phase extraction columns. The limit of detection of terfenadine in plasma is 0.1 ng/ml and the intra-assay coefficient of variation at 1 ng/ml is 10.1%. Plasma concentrations of terfenadine measured in six normal subjects following a 120 mg oral dose are reported.

Details

Language :
English
ISSN :
1387-2273
Volume :
693
Issue :
2
Database :
MEDLINE
Journal :
Journal of chromatography. B, Biomedical sciences and applications
Publication Type :
Academic Journal
Accession number :
9210438
Full Text :
https://doi.org/10.1016/s0378-4347(97)00083-2