Back to Search Start Over

Simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma using fluorescence detection.

Authors :
Peh KK
Yuen KH
Source :
Journal of chromatography. B, Biomedical sciences and applications [J Chromatogr B Biomed Sci Appl] 1997 May 23; Vol. 693 (1), pp. 241-4.
Publication Year :
1997

Abstract

A simple high-performance liquid chromatographic method using fluorescence detection was developed for the determination of acyclovir in human plasma. The method entailed direct injection of the plasma sample after deproteination. It is both specific and sensitive with a detection limit of 30 ng/ml at a signal-to-noise ratio of 3:1, and is thus suitable for use in pharmacokinetic studies of acyclovir. The method had a mean absolute recovery of 96%, while the within-day and between-day coefficients of variation and percentages error were all less than 8%. The calibration curve was linear over a concentration range of 62.5-4000 ng/ml.

Details

Language :
English
ISSN :
1387-2273
Volume :
693
Issue :
1
Database :
MEDLINE
Journal :
Journal of chromatography. B, Biomedical sciences and applications
Publication Type :
Academic Journal
Accession number :
9200543
Full Text :
https://doi.org/10.1016/s0378-4347(97)00041-8