[Intrabronchial migration and extraction of a fragment of an Accufix J-shaped atrial pacing catheter].

Authors :
Lefez C
Reybet-Degat O
Wolf JE
Louis P
Favre JP
David M
Source :
Archives des maladies du coeur et des vaisseaux [Arch Mal Coeur Vaiss] 1997 Feb; Vol. 90 (2), pp. 291-3.
Publication Year :
1997

Abstract

The authors report a case of exteriorisation and migration of a fragment of a J-shaped atrial Accufix Telectronix (Stimarec Class IV) responsible for a bronchial penetration without clinical repercussion. A systematic control chest X-ray detected this complication. The position of the metallic fragment was determined by chest CT scan and bronchial fibroscopy between the inferior right lobar bronches and artery confirming bronchial perforation. Rigid bronchoscopy with direct visualisation enabled extraction of the foreign body without complications.