Back to Search
Start Over
Tensile force of pacing lead extraction. A comparison between tined type and screw-in type.
- Source :
-
Japanese heart journal [Jpn Heart J] 1996 Jul; Vol. 37 (4), pp. 495-501. - Publication Year :
- 1996
-
Abstract
- Three months after implantation of endocardial pacing leads in 10 mongrel dogs, the tensile forces needed to remove 5 tined type and 5 screw-in type leads were compared experimentally. Mean maximum tensile force was 291 +/- 174 g with the screw-in type and 1174 +/- 369 g with the tined type (p < 0.01). Histopathologically, no changes suggesting any myocardial damage due to lead extraction were recognized in cases where screw-in type leads were used. These results support the clinical feasibility of safe and easy manual retraction of infected pacing leads of the screw-in type.
Details
- Language :
- English
- ISSN :
- 0021-4868
- Volume :
- 37
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Japanese heart journal
- Publication Type :
- Academic Journal
- Accession number :
- 8890763
- Full Text :
- https://doi.org/10.1536/ihj.37.495