Tensile force of pacing lead extraction. A comparison between tined type and screw-in type.

Authors :
Amitani S
Sohara H
Kurose M
Miyahara K
Misumi K
Sakamoto H
Source :
Japanese heart journal [Jpn Heart J] 1996 Jul; Vol. 37 (4), pp. 495-501.
Publication Year :
1996

Abstract

Three months after implantation of endocardial pacing leads in 10 mongrel dogs, the tensile forces needed to remove 5 tined type and 5 screw-in type leads were compared experimentally. Mean maximum tensile force was 291 +/- 174 g with the screw-in type and 1174 +/- 369 g with the tined type (p < 0.01). Histopathologically, no changes suggesting any myocardial damage due to lead extraction were recognized in cases where screw-in type leads were used. These results support the clinical feasibility of safe and easy manual retraction of infected pacing leads of the screw-in type.