Starch, gloves and extradural catheters.

Authors :
Green MA
Lam Y
Moss RF
Source :
British journal of anaesthesia [Br J Anaesth] 1995 Dec; Vol. 75 (6), pp. 768-70.
Publication Year :
1995

Abstract

We have investigated contamination of extradural catheters during normal handling with starch powdered gloves. In the laboratory, extradural catheters were handled in sterile fields with both powdered and non-powdered gloves, simulating preparation for patient insertion. The catheters together with glove samples were then examined using a Zeiss 940 scanning electron microscope. Microscopy of samples revealed starch contamination of the catheters handled with powdered gloves, especially in the side hole areas. We conclude that extradural catheters may be contaminated easily by starch powder from surgical gloves. This powder may then be deposited into the extradural space. The effect of starch in the extradural space is not known but starch is known to cause inflammatory and granulomatous reactions in other parts of the body as well as being directly allergenic. Powder contamination of catheters may be avoided easily by the use of powder-free gloves and we feel that these should be used whenever possible.