Back to Search
Start Over
Investigating the reliability and validity of the infrared tympanic thermometer: a learning experience.
- Source :
-
Journal of post anesthesia nursing [J Post Anesth Nurs] 1995 Oct; Vol. 10 (5), pp. 280-5. - Publication Year :
- 1995
-
Abstract
- This series of articles to date has discussed points that readers should be looking for when reading and critiquing a researcher article. This article presents a brief report of a research project investigating the reliability and validity of the infrared tympanic thermometer (ITT) in afebrile subjects. Using four subjects and 40 temperature diads, the ability of the ITT to predict the subjects' temperatures obtained with a mercury thermometer was very poor (regression coefficient = .39). Test-retest reliability using the same data set resulted in an equally poor correlation coefficient (r = .19). A critique of the methodology of the study is provided as a learning experience.