Back to Search Start Over

Comparison of convolution and ray-tracing methods for computing small blood vessel images in angiography.

Authors :
Koi K
Rossmann K
Source :
Medical physics [Med Phys] 1977 Mar-Apr; Vol. 4 (2), pp. 154-6.
Publication Year :
1977

Abstract

Two methods for computing x-ray images of small blood vessels in angiography are presently available, namely, convolution and ray tracing. The convolution method, which is simpler and more powerful than the ray-tracing method, is based on the assumption that blood vessel imaging is isoplantic, whereas ray tracing is considered to provide correct images. In this study, the approximation error (difference between two images, normalized by the maximum value) due to nonisoplanatic imaging was determined by computation of blood vessel images according to both methods. The approximation error for geometric conditions normally encountered in angiography was less than 0.01. It is concluded that an approximation error of this magnitude is negligible and that the convolution method can be applied instead of the ray-tracing method for the computation of images of small blood vessels.

Subjects

Subjects :
Angiography methods

Details

Language :
English
ISSN :
0094-2405
Volume :
4
Issue :
2
Database :
MEDLINE
Journal :
Medical physics
Publication Type :
Academic Journal
Accession number :
850511
Full Text :
https://doi.org/10.1118/1.594384