Back to Search
Start Over
Micro liquid chromatography-mass spectrometry with direct liquid introduction used for separation and quantitation of all-trans- and 13-cis-retinoic acids and their 4-oxo metabolites in human plasma.
- Source :
-
Journal of chromatography [J Chromatogr] 1993 Jul 23; Vol. 617 (1), pp. 129-35. - Publication Year :
- 1993
-
Abstract
- The separation and quantitation of the pentafluorobenzyl derivatives of all-trans- and 13-cis-retinoic acids and their 4-oxo metabolites in human plasma on micro high-performance liquid chromatographic columns (0.32 mm I.D.) is described. The column outlet was directly coupled to the source of a quadrupole mass spectrometer via a simple SFC-frit interface. Negative ion chemical ionization conditions were obtained by coaxial introduction of ammonia as a reagent gas. A signal-to-noise ratio well above 3 was obtained for 1 pg of each analyte injected. The limit of quantitation calculated from spiked biological plasma extracts was 0.3 ng/ml.
Details
- Language :
- English
- Volume :
- 617
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of chromatography
- Publication Type :
- Academic Journal
- Accession number :
- 8376526
- Full Text :
- https://doi.org/10.1016/0378-4347(93)80431-3