Back to Search Start Over

Column liquid chromatographic analysis of quinine in human plasma, saliva and urine.

Authors :
Babalola CP
Bolaji OO
Dixon PA
Ogunbona FA
Source :
Journal of chromatography [J Chromatogr] 1993 Jun 23; Vol. 616 (1), pp. 151-4.
Publication Year :
1993

Abstract

A new simple, selective and reproducible high-performance liquid chromatographic method for the determination of quinine in plasma, saliva and urine is described. The ion-pair method was carried out on a reversed-phase C18 column, using perchlorate ion as the counter ion and ultraviolet detection at 254 nm. Quinine was well resolved from its major metabolite, 3-hydroxyquinine, and the internal standard, primaquine. The limit of detection was 10 ng/ml and the recovery was greater than 90% from the three biological fluids.

Details

Language :
English
Volume :
616
Issue :
1
Database :
MEDLINE
Journal :
Journal of chromatography
Publication Type :
Academic Journal
Accession number :
8376486
Full Text :
https://doi.org/10.1016/0378-4347(93)80482-j