Decreased signal in Emit assays of drugs of abuse in urine after ingestion of aspirin: potential for false-negative results.
- Source :
-
Clinical chemistry [Clin Chem] 1994 Apr; Vol. 40 (4), pp. 608-12. - Publication Year :
- 1994
-
Abstract
- During routine drug analysis with the Syva d.a.u. Emit immunoassays we observed a high frequency of urines with lower rates of changes in absorbance (delta A R) than the rate for a drug-free urine calibrator. Many of these urines contained salicylates. Among 40 urines with apparent salicylate concentrations between 15 and 420 mg/dL tested for benzoylecgonine (BE), 20 had delta A R < -4 (range +2 to -28 mA/min). The rates decreased with increasing salicylate: delta A R = -0.057 x (salicylate, mg/dL) -0.22 mA/min (r = 0.85, n = 40, P < 0.01). Urines from 100 control subjects (no salicylate) had mean +/- SD delta A R values of -1.05 +/- 2.2 mA/min (range +3 to -7; only two were < -4 mA/min). Although direct addition of salicylic acid (200 mg/dL) to urine specimens did not reproduce the negative bias, ingestion of aspirin (acetylsalicylic acid) did by -0.09 mA/min per 1 mg/dL (72.4 mumol/L) salicylate. Negative biases observed for other Emit d.a.u. assays after salicylate ingestion lead us to conclude that ingestion of therapeutic doses of aspirin may cause false-negative results for drug screens in urines by this technology.