Back to Search
Start Over
Quality assessment in in vivo NMR spectroscopy: V. Multicentre evaluation of prototype test objects and protocols for performance assessment in small bore MRS equipment.
- Source :
-
Magnetic resonance imaging [Magn Reson Imaging] 1995; Vol. 13 (1), pp. 159-67. - Publication Year :
- 1995
-
Abstract
- This paper reports the results of multicentre studies aimed at designing, constructing, and evaluating prototype test objects for performance assessment in small-bore MRS systems, by utilizing the test protocols already proposed by the EEC COMAC-BME Concerted Action for clinical MRS equipment. Three classes of test objects were considered: (1) a multicompartment test object for 31P MRS measurements performed with slice-selective sequences; (2) a two-compartment test object for volume-selection 1H MRS; and (3) two-compartment test objects for assessing the performance of experimental systems using ISIS as volume localization sequence in 31P MRS. The results suggested the interest of adopting some of these prototypes for improving the comparison of spectroscopy data obtained from different sites, for providing useful means of quality assurance in experimental MRS, and facilitating the validation of new localization sequences.
- Subjects :
- Quality Control
Magnetic Resonance Spectroscopy
Subjects
Details
- Language :
- English
- ISSN :
- 0730-725X
- Volume :
- 13
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Magnetic resonance imaging
- Publication Type :
- Academic Journal
- Accession number :
- 7898275
- Full Text :
- https://doi.org/10.1016/0730-725x(94)00091-g