Back to Search Start Over

Simultaneous determination of vitamins D3, E and K1 and retinyl palmitate in cattle plasma by liquid chromatography with a narrow-bore column.

Authors :
Gomis DB
Escotet Arias VJ
Fidalgo Alvarez LE
GutiƩrrez Alvarez MD
Source :
Journal of chromatography. B, Biomedical applications [J Chromatogr B Biomed Appl] 1994 Oct 03; Vol. 660 (1), pp. 49-55.
Publication Year :
1994

Abstract

A reversed-phase high performance liquid chromatographic method is described for the simultaneous determination of vitamins D3, E and K1 and retinyl palmitate in plasma. Narrow-bore columns are recommended because this alternative provides a good separation efficiency, plus greater economy and sensitivity. Detection limits for individual vitamins range from 0.42 to 2.8 ng. All vitamins were separated in less than 9 min. Recovery studies showed good results for all solutes (88.8-100.3%) and the intra-day coefficients of variations ranged from 1.0 to 4.5%. This method permits the simple determination of fat-soluble vitamins using 1 ml of cattle heparinized plasma.

Details

Language :
English
ISSN :
1572-6495
Volume :
660
Issue :
1
Database :
MEDLINE
Journal :
Journal of chromatography. B, Biomedical applications
Publication Type :
Academic Journal
Accession number :
7858723
Full Text :
https://doi.org/10.1016/0378-4347(94)00279-7