Bitewing-based alveolar bone densitometry: digital imaging resolution requirements.

Authors :
Hildebolt CF
Bartlett TQ
Brunsden BS
Hente NL
Gravier MJ
Walkup RK
Shrout MK
Vannier MW
Source :
Dento maxillo facial radiology [Dentomaxillofac Radiol] 1994 Aug; Vol. 23 (3), pp. 129-34.
Publication Year :
1994

Abstract

The resolution requirements were determined for detection of incremental alveolar bone loss from digitized Ektaspeed radiographs. Ten clinical radiographs were examined with a calibrated optical microscope to measure the smallest feature of interest discernible for alveolar bone. Images of trabeculae > 100 microns in diameter could be identified, but smaller ones could not be resolved. The Nyquist Criterion requires sampling with 50 microns (or smaller) pixels to measure such features. Numerous 25 microns Ag aggregates were present. Fifty microns resolution is a practical compromise between noise level and feature resolution. In another example of 10 bitewings digitized at 8 bit grey scale, about half the available range was used for alveolar bone, resulting in uncertainty, over a range of 2 optical density (OD) units, of about 1.42% at the average OD for alveolar bone (1.1). With the same radiographs digitized at 12 bit resolution, over 2000 of 4096 grey scales were used with a corresponding uncertainty of about 0.09%. Twelve-bit precision was also able to resolve smaller increments in an aluminium stepwedge than was 8 bit precision. Twelve-bit grey scale and 50 microns spatial resolution are recommended for alveolar bone densitometry performed with Ektaspeed radiographs.

Details

Language :
English
ISSN :
0250-832X
Volume :
23
Issue :
3
Database :
MEDLINE
Journal :
Dento maxillo facial radiology
Publication Type :
Academic Journal
Accession number :
7835512
Full Text :
https://doi.org/10.1259/dmfr.23.3.7835512