Back to Search Start Over

Performance of the Coumatrak system in a large anticoagulation clinic.

Authors :
Foulis PR
Wallach PM
Adelman HM
Sanford BH
McCain J
Reed D
Schlede CM
Kokseng CU
Taylor CD
Source :
American journal of clinical pathology [Am J Clin Pathol] 1995 Jan; Vol. 103 (1), pp. 98-102.
Publication Year :
1995

Abstract

There is a need to monitor anticoagulation accurately, inexpensively, and rapidly. The accuracy and precision of a simple fingerstick method was studied in a large outpatient anticoagulation clinic using the Coumatrak method. The Coumatrak apparatus has been studied in the home setting, and three recent reports suggest that it is practical, accurate, and possibly superior to the standard method. These results differ from recently published studies. This technique was found to be less than acceptable in precision and accuracy. This method requires further study before it can be recommended for wide-spread use in making decisions for patient care.

Details

Language :
English
ISSN :
0002-9173
Volume :
103
Issue :
1
Database :
MEDLINE
Journal :
American journal of clinical pathology
Publication Type :
Academic Journal
Accession number :
7817953
Full Text :
https://doi.org/10.1093/ajcp/103.1.98