Back to Search Start Over

An assessment of the Corning electrophoretic procedure for high density lipoprotein cholesterol estimation.

Authors :
Bullock DG
Carter TJ
Cosgrove CM
Hughes SV
Source :
Annals of clinical biochemistry [Ann Clin Biochem] 1980 May; Vol. 17 (3), pp. 148-52.
Publication Year :
1980

Abstract

The electrophoretic procedure for high density lipoprotein cholesterol determination recently introduced by Corning Medical has been assessed. The results corrrelated well with those obtained by phosphotungstate/Mg precipitation and ultracentrifuge procedures, but the between-batch CV was 6% in the clinically important range below 1 mmol/l. The densitometer, which must be carefully aligned before use for this analysis, seemed sensitive to variations in the plates and patterns produced. The technique required that great care, especially with timing, and appeared expensive.

Details

Language :
English
ISSN :
0004-5632
Volume :
17
Issue :
3
Database :
MEDLINE
Journal :
Annals of clinical biochemistry
Publication Type :
Academic Journal
Accession number :
7406438
Full Text :
https://doi.org/10.1177/000456328001700309