Back to Search Start Over

Calibration of a 238Pu,Be facility for partial-body measurements of organ cadmium.

Authors :
Morgan WD
Ellis KJ
Vartsky D
Yasumura S
Cohn SH
Source :
Physics in medicine and biology [Phys Med Biol] 1981 Jul; Vol. 26 (4), pp. 577-90.
Publication Year :
1981

Abstract

An improved instrument is described for the measurement of liver and kidney cadmium by in vivo neutron activation analysis in both occupationally and environmentally exposed persons. Detailed calibrations of the instrument used in a study of 83 male workers at a cadmium production plant are give. The importance of accurate organ localisation by ultrasound is stressed, without which errors of 40 and 25% in individual and group kidney measurements, respectively, can occur. The detection limit (2 SD of the background) is 2.2 mg cadmium in the kidney and 1.5 micrograms g-1 (wet weight) in the liver for a local dose of 4.7 mSv. This instrument therefore combines the advantages of portability with high sensitivity of detection of cadmium.

Details

Language :
English
ISSN :
0031-9155
Volume :
26
Issue :
4
Database :
MEDLINE
Journal :
Physics in medicine and biology
Publication Type :
Academic Journal
Accession number :
7255543
Full Text :
https://doi.org/10.1088/0031-9155/26/4/002