Back to Search Start Over

Control system for circulatory assist devices: determination of suitable control variables.

Authors :
Puri NN
Li JK
Fich S
Welkowitz W
Source :
Transactions - American Society for Artificial Internal Organs [Trans Am Soc Artif Intern Organs] 1982; Vol. 28, pp. 127-32.
Publication Year :
1982

Abstract

This paper studies the problem of identification of suitable biological variables for optimal effectiveness of a CAD. To identify suitable error signals, a closed loop simulation and analysis of the assisted circulation was carried out. Results show that the combination of heart rate and left atrial pressure is very effective as an input error signal for either series or parallel assisted circulation when simulated heart failure is severe (greater than 50% decrease in left ventricular elastance). During simulated exercise (50% reduction in peripheral resistance) such a combination is even more effective. In addition, simulation results show that aortic pressure and cardiac output are not suitable control variables for CAD.

Details

Language :
English
ISSN :
0066-0078
Volume :
28
Database :
MEDLINE
Journal :
Transactions - American Society for Artificial Internal Organs
Publication Type :
Academic Journal
Accession number :
7164228