Back to Search Start Over

Electrode for sensing fluoride ion activity in solution.

Authors :
Frant MS
Ross JW Jr
Source :
Science (New York, N.Y.) [Science] 1966 Dec 23; Vol. 154 (3756), pp. 1553-5.
Publication Year :
1966

Abstract

Electrodes constructed from single-crystal sections of rare earth fluorides respond to fluoride ion activity over more than five orders of magnitude and show a high selectivity for fluoride over other common anions. These electrodes can be used for either direct measuremnent of fluoride ion activity or detection of the end point in titration.

Details

Language :
English
ISSN :
0036-8075
Volume :
154
Issue :
3756
Database :
MEDLINE
Journal :
Science (New York, N.Y.)
Publication Type :
Academic Journal
Accession number :
5924922
Full Text :
https://doi.org/10.1126/science.154.3756.1553