Back to Search Start Over

Automated high-performance liquid chromatographic determination of spiramycin by direct injection of plasma, using column-switching for sample clean-up.

Authors :
Dow J
Lemar M
Frydman A
Gaillot J
Source :
Journal of chromatography [J Chromatogr] 1985 Nov 08; Vol. 344, pp. 275-83.
Publication Year :
1985

Abstract

A fully automated high-performance liquid chromatographic method is described for the determination of spiramycin 1 in plasma. First, 1 ml of plasma is diluted with 1.5 ml of 4% acetonitrile containing spiramycin 2 as internal standard, and 1 ml is then injected via an automatic sampling unit. A pneumatic valve, which is remote-controlled by the programmable timer of an integrator, switches the sample, initially injected onto a precolumn for sample clean-up, to an analytical column for sample separation. This method was compared with a microbiological assay and has been successfully applied to pharmacokinetic studies on spiramycin in humans.

Details

Language :
English
Volume :
344
Database :
MEDLINE
Journal :
Journal of chromatography
Publication Type :
Academic Journal
Accession number :
4086548
Full Text :
https://doi.org/10.1016/s0378-4347(00)82028-9