Back to Search
Start Over
High-resolution cryo-EM using a common LaB 6 120-keV electron microscope equipped with a sub-200-keV direct electron detector.
- Source :
-
Science advances [Sci Adv] 2025 Jan 03; Vol. 11 (1), pp. eadr0438. Date of Electronic Publication: 2025 Jan 03. - Publication Year :
- 2025
-
Abstract
- High-resolution cryo-electron microscopy (cryo-EM) requires costly 200- to 300-keV cryo-transmission electron microscopes (cryo-TEMs) with field emission gun (FEG) sources, stable columns, constant-powered lenses, autoloader, and direct electron detectors (DED). Recent advances in 100-keV imaging with the emergence of sub-200-keV optimized DED technology promises the development of more affordable cryo-TEMs. So far, 100-keV imaging has required microscopes with FEG sources. We here explored whether a standard 120-keV TEMs with thermionic lanthanum hexaboride (LaB <subscript>6</subscript> ) source can be upgraded with a sub-200-keV DED for high-resolution cryo-EM. Using this imaging configuration, we successfully obtained a 2.65 Å reconstruction for apoferritin, 4.33 Å for 64-kDa hemoglobin, and 4.4 Å for an asymmetric 153kDa membrane protein GPCR. All results were achieved using standard automated data collection with SerialEM, demonstrating the feasibility to collect large cryo-EM datasets with a side-entry cryo-holder. These results showcase a widely accessible solution to obtaining interpretable cryo-EM structures at low cost and contribute to the "democratization" of cryo-EM.
Details
- Language :
- English
- ISSN :
- 2375-2548
- Volume :
- 11
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Science advances
- Publication Type :
- Academic Journal
- Accession number :
- 39752481
- Full Text :
- https://doi.org/10.1126/sciadv.adr0438