Back to Search Start Over

A selective dual-signal electrochemical paper-based device using imprinted sensors for voltammetric and impedance analysis of 4-NQO and carcinoembryonic antigen (CEA).

Authors :
Ngaosri P
Karuwan C
Wanram S
Jarujamrus P
Citterio D
Amatatongchai M
Source :
Analytica chimica acta [Anal Chim Acta] 2024 Nov 22; Vol. 1330, pp. 343273. Date of Electronic Publication: 2024 Sep 21.
Publication Year :
2024

Abstract

Background: Carcinoembryonic Antigen (CEA) and 4-nitroquinoline-N-oxide (4-NQO) are cancer markers that play a crucial role in tumor risk assessment and early cancer diagnosis. Therefore, it is in demand to develop a fast, accurate, simple, and cost-effective method to detect these cancer markers for quick and early stage-cancer diagnosis and treatment.<br />Results: Herein, we report a dual signaling approach for direct and indirect signal transduction of cancer biomarker binding on molecularly imprinted-electrodes integrated on an ePAD, enabling sensitive and selective quantitative analysis of 4-NQO and CEA. Multiple test zones on a single device based on artificial recognition sites using core-shell Cu-MOF@MIP were developed to create a selective "Dual-signal-ePAD". An effective and facile imprinting method on a Cu-MOF surface based on the synthesized of 4-VPBA/MAA/TRIM polymer. It resulted in a novel MIP useable as a high-performance sensing tool for the ultrasensitive and specific quantification of the target cancer biomarkers. Multiple tests consisted of voltammetric analysis of 4-NQO via oxidation of the analyte on a WE <subscript>1</subscript> , and impedance analysis of CEA via the molecular interaction with the imprinted WE <subscript>2</subscript> .<br />Significance: The proposed device provides enhanced electrochemical signal amplification due to its high conductivity, electrocatalytic activity, and the target-specific recognition sites for accurate determination of 4-NQO and CEA targets. The Dual-signal-ePAD exhibits good voltammetric and EIS response, oxidation currents of 4-NQO at -0.06 V and R <subscript>CT</subscript> of CEA, resulting in high selectivity and sensitivity with a linear response range covering the concentration range of clinical interest for both analytes. Dual-signal ePAD is a good candidate for clinical screening and POCT.<br />Competing Interests: Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (Copyright © 2024 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-4324
Volume :
1330
Database :
MEDLINE
Journal :
Analytica chimica acta
Publication Type :
Academic Journal
Accession number :
39489955
Full Text :
https://doi.org/10.1016/j.aca.2024.343273