Back to Search
Start Over
Time course of visual attention in rats by atomic magnetometer.
- Source :
-
PloS one [PLoS One] 2024 Oct 29; Vol. 19 (10), pp. e0312589. Date of Electronic Publication: 2024 Oct 29 (Print Publication: 2024). - Publication Year :
- 2024
-
Abstract
- Atomic magnetometer (AM) is utilized to non-invasively detect event-related magnetic fields (ERMFs) evoked by visual stimuli in rats. The aim of this study was to investigate the relationship between N2-like amplitude and visual attention. To achieve this, we combined the AM with a visual stimulation system and employed the passive single-stimulus paradigm. By measuring the ERMFs at various inter-stimulus intervals (ISIs) with a sensitivity of 20 fT/[Formula: see text], we analyzed the effects of the ISI and the 'habituation' resulting from repeated stimuli on the N2-like amplitude. Our method serves as a valuable reference for studying the passive single-stimulus paradigm and the time course of mammalian attention.<br />Competing Interests: The authors have declared that no competing interests exist.<br /> (Copyright: © 2024 Liu et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.)
Details
- Language :
- English
- ISSN :
- 1932-6203
- Volume :
- 19
- Issue :
- 10
- Database :
- MEDLINE
- Journal :
- PloS one
- Publication Type :
- Academic Journal
- Accession number :
- 39471149
- Full Text :
- https://doi.org/10.1371/journal.pone.0312589