Evaluation of the Seegene Allplex™ RV master assay for one-step simultaneous detection of eight respiratory viruses in nasopharyngeal specimens.
- Source :
-
Journal of virological methods [J Virol Methods] 2025 Jan; Vol. 331, pp. 115042. Date of Electronic Publication: 2024 Oct 09. - Publication Year :
- 2025
-
Abstract
- Background: The Seegene Allplex™ RV Master (RVM) assay is a one-step multiplex real-time reverse transcription polymerase chain reaction (RT-PCR) system for detecting eight viral respiratory pathogens from nasopharyngeal swab, aspirate, and bronchoalveolar lavage specimens. The eight RVM targets are: severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), Influenza A (Flu A), Influenza B (Flu B), Human respiratory syncytial virus (RSV), adenovirus (AdV), rhinovirus (HRV), parainfluenza virus (PIV), and metapneumovirus (MPV). The assay is based on Seegene's multiple detection temperature (MuDT) technology and provides cycle threshold (Ct) values for each of its viral targets upon PCR completion.<br />Objective: We aimed to evaluate the diagnostic performance of the RVM assay by calculating sensitivity, specificity, accuracy, Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Percent Agreement (PPA), Negative Percent Agreement (NPA), and Overall Percent Agreement (OPA) compared to definite diagnosis and analogous reference assays.<br />Study Design: Diagnostic sensitivity, specificity, accuracy, PPV, and NPV were calculated by comparing the results of the RVM assay to that of definite diagnosis assays; while PPA, NPA, and OPA were calculated by comparing results of the RVM assay to that of analogous reference products. Definite diagnosis and reference methods comprised whole genome sequencing and PCR genotyping, the Allplex™ SARS-CoV-2/FluA/FluB/RSV and Respiratory Panels 1, 2, and 3 assays (Seegene), and the Xpert® Xpress SARS-CoV-2/FluA/FluB/RSV Plus assay (Cepheid). Reproducibility of the RVM assay using fully-automated and semi-automated nucleic acid (NA) extraction workflows and as performed by independent operators was also assessed. In total, 249 positive respiratory specimens and at least 50 negative specimens for each target tested were used for this evaluation study.<br />Results: Sensitivity, specificity, accuracy, PPV, NPV, PPA, NPA, and OPA ranged from 95.7 % to 100 % for detecting all eight targets tested on the RVM assay. Reproducibility PPA, NPA, and OPA between automated and semi-automated NA extraction workflows were all >97.9 %, while the reproducibility PPA, NPA and OPA between independent operators were all 100 %.<br />Conclusion: These results demonstrate a high level of sensitivity, specificity, accuracy and diagnostic predictive value of the RVM assay and high agreement with comparable reference assays in identifying all eight of its targets. Taken together, our study underscores the diagnostic utility of the RVM assay in detecting eight viral respiratory pathogens.<br />Competing Interests: Declaration of Competing Interest None of the authors of this manuscript have any conflicts of interest to declare.<br /> (Copyright © 2024 The Authors. Published by Elsevier B.V. All rights reserved.)
- Subjects :
- Humans
Adult
Middle Aged
Child
Child, Preschool
Female
Aged
Adolescent
Infant
Male
Young Adult
Viruses isolation & purification
Viruses genetics
Viruses classification
SARS-CoV-2 genetics
SARS-CoV-2 isolation & purification
Real-Time Polymerase Chain Reaction methods
COVID-19 diagnosis
COVID-19 virology
Molecular Diagnostic Techniques methods
Nasopharynx virology
Sensitivity and Specificity
Respiratory Tract Infections virology
Respiratory Tract Infections diagnosis
Multiplex Polymerase Chain Reaction methods
Subjects
Details
- Language :
- English
- ISSN :
- 1879-0984
- Volume :
- 331
- Database :
- MEDLINE
- Journal :
- Journal of virological methods
- Publication Type :
- Academic Journal
- Accession number :
- 39384158
- Full Text :
- https://doi.org/10.1016/j.jviromet.2024.115042