Back to Search Start Over

Transcatheter Non-Acute Retrieval of the Tines-Based Leadless Ventricular Pacemaker.

Authors :
Funasako M
Hála P
Janotka M
Šorf J
Machová L
Petrů J
Chovanec M
Škoda J
Šedivá L
Šimon J
Dujka L
Reddy VY
Neužil P
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2024 Oct 07. Date of Electronic Publication: 2024 Oct 07.
Publication Year :
2024
Publisher :
Ahead of Print

Abstract

Objectives: We report our single-center experience of mid-term to long-term retrieval and reimplantation of a tines-based leadless pacemaker (Micra transcatheter pacing system; TPS).<br />Backgrounds: The TPS is a clinically effective alternative to transvenous single-chamber ventricular pacemakers. Whereas it is currently recommended to abandon the TPS at the end of device life, catheter-based retrieval may be favorable in specific scenarios.<br />Methods: We report on nine consecutive patients with the implanted TPS who subsequently underwent transcatheter retrieval attempts. The retrieval system consists of the original TPS delivery catheter and an off-the-shelf single-loop 7 mm snare. The procedure was guided by fluoroscopy and intracardiac echocardiography.<br />Results: After an implantation duration of 3.1 ± 2.8 years (range 0.4-9.0), the overall retrieval success rate was 88.9% (8 of 9 patients). The mean procedure time was 89 ± 16 minutes, and fluoroscopy time was 18.0 ± 6.6 min. No procedure-related adverse device events occurred. In the one unsuccessful retrieval, intracardiac echocardiography revealed that the TPS was partially embedded in the ventricular tissue surrounding the leadless pacemaker body in the right ventricle. After retrieval, three patients were reimplanted with a new TPS device. All implantations were successful without complications.<br />Conclusions: A series of transvenous late retrievals of implanted TPS devices demonstrated safety and feasibility, followed by elective replacement with a new leadless pacing device or conventional transvenous pacing system. This provides a viable end-of-life management alternative to simple abandonment of this leadless pacemaker.<br /> (© The Author(s) 2024. Published by Oxford University Press on behalf of the European Society of Cardiology.)

Details

Language :
English
ISSN :
1532-2092
Database :
MEDLINE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
39374848
Full Text :
https://doi.org/10.1093/europace/euae256