Fraction of reverse impedance change (FRIC): a quantitative electrical impedance tomography measure of intrapulmonary pendelluft.

Authors :
Adler A
Becher T
Händel C
Frerichs I
Source :
Physiological measurement [Physiol Meas] 2024 Oct 28; Vol. 45 (10). Date of Electronic Publication: 2024 Oct 28.
Publication Year :
2024

Abstract

Objective . Pendelluft is the movement of air between lung regions, and electrical impedance tomography (EIT) has shown an ability to detect and monitor it. Approach. In this note, we propose a functional EIT measure which quantifies the reverse airflow seen in pendelluft: the Fraction of Reverse Impedance Change (FRIC). Main Results . FRIC measures the fraction of reverse flow in each pixel waveform (as an image) or globally (as a single parameter). Significance . Such a measure is designed to be a more specific measure than previous approaches, to enable comparative studies of the pendelluft, and to help clarify the effect of ventilation strategies.<br /> (Creative Commons Attribution license.)