Back to Search Start Over

Turbolysis: A low-cost, small footprint alternative to commercial bead beaters for cell lysis.

Authors :
Limberis JD
Metcalfe JZ
Source :
HardwareX [HardwareX] 2024 Aug 27; Vol. 19, pp. e00576. Date of Electronic Publication: 2024 Aug 27 (Print Publication: 2024).
Publication Year :
2024

Abstract

turboLysis is a novel mechanical cell lysis device that utilizes small beads to efficiently lyse tough cells like Mycobacterium, Saccharomyces, and Arabidopsis. We compared turboLysis to bead beating using the BeadBug 6 for several concentrations of Mycobacterium tuberculosis roughly correlated to the bacterial load commonly seen in patient samples. turboLysis performed similarly to the BeadBug at low bacterial concentrations and outperformed it at high concentrations above 2x10 <superscript>5</superscript> CFU/ml (p < 0.005). Thus, turboLysis offers good cell lytic performance in a small form factor at a low cost.<br />Competing Interests: The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.<br /> (© 2024 The Author(s).)

Details

Language :
English
ISSN :
2468-0672
Volume :
19
Database :
MEDLINE
Journal :
HardwareX
Publication Type :
Academic Journal
Accession number :
39280234
Full Text :