Back to Search Start Over

External quality assessment performance in ten countries: an IFCC global laboratory quality project.

Authors :
Bais R
Vassault A
Blasutig IM
Dabla PK
Lin J
Perret-Liaudet A
Thomas A
Cendejas KA
Wheeler SE
Giannoli JM
Meng QH
Amann EP
Source :
Clinical chemistry and laboratory medicine [Clin Chem Lab Med] 2024 Sep 02; Vol. 62 (12), pp. 2435-2443. Date of Electronic Publication: 2024 Sep 02 (Print Publication: 2024).
Publication Year :
2024

Abstract

Objectives: This study aimed to assess the validity of external quality assessment (EQA) laboratory results across various cultural and environmental contexts and to identify potential improvement areas.<br />Methods: The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Task Force on Global Laboratory Quality (TF-GLQ) conducted a 2-year study (2022 and 2023) in which EQA materials, related software and online training was provided by a commercial vendor to 100 laboratories in ten IFCC member society countries. The results were analysed on a monthly basis by the TF-GLQ, to show the number of submissions per country, tests per lab, acceptability rates, random failures and to get a measure of which analytes performed poorly.<br />Results: The EQA material was dispatched on a quarterly basis. Some countries had problems with customs releasing the material in a timely manner, resulting in laboratories not receiving them on time leading to no submission. We report here the results for the second year of the survey. The number of examinations varied between laboratories, ranging from seven to 84 analytes. Of the ten countries surveyed, six averaged greater than 90 % acceptable results over the whole 12-months cycle, one had unacceptable results for two of the nine months they returned results and the other four were considered to not perform to an acceptable standard.<br />Conclusions: All 100 participating laboratories indicated satisfaction with the EQA survey and related services, including on-site training, and report handling. However, specimen receiving issues, suggest benefits in dispatching materials for a full 12-month cycle. Significant discrepancies in EQA performance indicate that four countries require long-term assistance, training and guidance. To ensure reliable patient results, promoting EQA in certain countries is essential to achieve the required level of quality.<br /> (© 2024 Walter de Gruyter GmbH, Berlin/Boston.)

Details

Language :
English
ISSN :
1437-4331
Volume :
62
Issue :
12
Database :
MEDLINE
Journal :
Clinical chemistry and laboratory medicine
Publication Type :
Academic Journal
Accession number :
39238208
Full Text :
https://doi.org/10.1515/cclm-2024-0859