Back to Search Start Over

Assessment of Subject Head Motion in Diffusion MRI.

Authors :
Topolnjak E
Gao C
Beason-Held LL
Resnick SM
Schilling KG
Landman BA
Source :
Proceedings of SPIE--the International Society for Optical Engineering [Proc SPIE Int Soc Opt Eng] 2024 Feb; Vol. 12926. Date of Electronic Publication: 2024 Apr 02.
Publication Year :
2024

Abstract

Subject head motion during the acquisition of diffusion-weighted imaging (DWI) of the brain induces artifacts and affects image quality. Information about the frequency and extent of motion could reveal which aspects of motion correction are most necessary. Therefore, we investigate the extent of translation and rotation among participants, and how the motion changes during the scan acquisition. We analyze 5,380 DWI scans from 1,034 participants. We measure the rotations and translations in the sagittal, coronal and transverse planes needed to align the volumes to the first and previous volumes, as well as the displacement. The different types of motion are compared with each other and compared over time. The largest rotation (per minute) is around the right - left axis (median 0.378 °/min, range 0.000 - 11.466°) and the largest translation (per minute) is along the anterior - posterior axis (median 1.867 mm/min, range 0.000 - 10.944 mm). We additionally observe that spikes in movement occur at the beginning of the scan, particularly in anterior - posterior translation. The results show that all scans are affected by subtle head motion, which may impact subsequent image analysis.

Details

Language :
English
ISSN :
0277-786X
Volume :
12926
Database :
MEDLINE
Journal :
Proceedings of SPIE--the International Society for Optical Engineering
Publication Type :
Academic Journal
Accession number :
39220213
Full Text :
https://doi.org/10.1117/12.3006633