Back to Search Start Over

Long-Term Plate Complications in Patient-Specific Plates Utilizing Computer-Aided Design.

Authors :
Yang S
Morton Z
Sautter M
Young G
Petrisor D
Chandra SR
Wax MK
Source :
The Laryngoscope [Laryngoscope] 2024 Dec; Vol. 134 (12), pp. 4929-4934. Date of Electronic Publication: 2024 Jul 17.
Publication Year :
2024

Abstract

Objective: Assess the long-term plate complications with patient-specific plates (PSPs) created with computer-aided design (CAD) and computer-aided manufacturing (CAM) for fibula free flap reconstructions for mandibular defects.<br />Methods: Retrospective chart review from January 2010 to July 2022 of patients who underwent mandibular reconstruction with a fibula free flap and PSP. Primary outcome was plating-related complications, defined as plate exposure, fracture, loose screws, and plate removal.<br />Results: A total of 221 patients underwent PSP fibula reconstruction. Average age was 59.8 + 14.3 years old with male to female ratio of 2:1. Squamous cell carcinoma of the mandible was the most common reason for resection, 47.5%, n = 105. Plate removal occurred in 11% of patients (n = 25) about 17.4 months after the initial surgery. Plates were removed due to exposure (76%, n = 19) or screw loosening (24%, n = 6). Malignancy was associated with an increased risk of plate complications when compared to benign tumor (odds ratio [OR] 9.04, confidence interval [CI] 1.36-3.85), osteonecrosis (OR 1.38, CI 0.59-3.48), and trauma (OR 1.26, CI 0.23-12.8). Postoperative radiation therapy (OR 2.27, CI 1.07-4.82, p = 0.026) and surgical site infection (OR 9.22, CI 4.11-21.88, p = 0.001) were associated with more plate complications.<br />Conclusions: CAD creates PSPs that remain stable in the majority of patients over the long term. Plate removal is less compared to non-PSP reconstruction. Consideration of the soft tissue envelope over the plate and management of perioperative infection at the time of surgery should be entertained.<br />Level of Evidence: 4 Laryngoscope, 134:4929-4934, 2024.<br /> (© 2024 The American Laryngological, Rhinological and Otological Society, Inc.)

Details

Language :
English
ISSN :
1531-4995
Volume :
134
Issue :
12
Database :
MEDLINE
Journal :
The Laryngoscope
Publication Type :
Academic Journal
Accession number :
39016186
Full Text :
https://doi.org/10.1002/lary.31640