Clinical outcomes and predictors of transapical transcatheter mitral valve replacement: the Tendyne Expanded Clinical Study.

Authors :
Conradi L
Ludwig S
Sorajja P
Duncan A
Bethea B
Dahle G
Babaliaros V
Guerrero M
Thourani V
Dumonteil N
Modine T
Garatti A
Denti P
Leipsic J
Chuang ML
Blanke P
Muller DW
Badhwar V
Source :
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology [EuroIntervention] 2024 Jul 15; Vol. 20 (14), pp. e887-e897. Date of Electronic Publication: 2024 Jul 15.
Publication Year :
2024

Abstract

Background: Transcatheter mitral valve replacement (TMVR) is a therapeutic option for patients with severe mitral regurgitation (MR) who are ineligible for conventional surgery. There are limited data on the outcomes of large patient cohorts treated with TMVR.<br />Aims: This study aimed to investigate the outcomes and predictors of mortality for patients treated with transapical TMVR.<br />Methods: This analysis represents the clinical experience of all patients enrolled in the Tendyne Expanded Clinical Study. Patients with symptomatic MR underwent transapical TMVR with the Tendyne system between November 2014 and June 2020. Outcomes and adverse events up to 2 years, as well as predictors of short-term mortality, were assessed.<br />Results: A total of 191 patients were treated (74.1±8.0 years, 62.8% male, Society of Thoracic Surgeons Predicted Risk of Mortality 7.7±6.6%). Technical success was achieved in 96.9% (185/191), and there were no intraprocedural deaths. At 30-day, 1- and 2-year follow-up, the rates of all-cause mortality were 7.9%, 30.8% and 40.5%, respectively. Complete MR elimination (MR <1+) was observed in 99.3%, 99.1% and 96.3% of patients, respectively. TMVR treatment resulted in consistent improvement of New York Heart Association Functional Class and quality of life up to 2 years (both p<0.001). Independent predictors of early mortality were age (odds ratio [OR] 1.11; p=0.003), pulmonary hypertension (OR 3.83; p=0.007), and institutional experience (OR 0.40; p=0.047).<br />Conclusions: This study investigated clinical outcomes in the full cohort of patients included in the Tendyne Expanded Clinical Study. The Tendyne TMVR system successfully eliminated MR with no intraprocedural deaths, resulting in an improvement in symptoms and quality of life. Continued refinement of clinical and echocardiographic risks will be important to optimise longitudinal outcomes.

Details

Language :
English
ISSN :
1969-6213
Volume :
20
Issue :
14
Database :
MEDLINE
Journal :
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
39007829
Full Text :
https://doi.org/10.4244/EIJ-D-23-00904