Back to Search
Start Over
Turbulent flow field in maglev centrifugal blood pumps of CH-VAD and HeartMate III: secondary flow and its effects on pump performance.
- Source :
-
Biomechanics and modeling in mechanobiology [Biomech Model Mechanobiol] 2024 Oct; Vol. 23 (5), pp. 1571-1589. Date of Electronic Publication: 2024 May 31. - Publication Year :
- 2024
-
Abstract
- Secondary flow path is one of the crucial aspects during the design of centrifugal blood pumps. Small clearance size increases stress level and blood damage, while large clearance size can improve blood washout and reduce stress level. Nonetheless, large clearance also leads to strong secondary flows, causing further blood damage. Maglev blood pumps rely on magnetic force to achieve rotor suspension and allow more design freedom of clearance size. This study aims to characterize turbulent flow field and secondary flow as well as its effects on the primary flow and pump performance, in two representative commercial maglev blood pumps of CH-VAD and HeartMate III, which feature distinct designs of secondary flow path. The narrow and long secondary flow path of CH-VAD resulted in low secondary flow rates and low disturbance to the primary flow. The flow loss and blood damage potential of the CH-VAD mainly occurred at the secondary flow path, as well as the blade clearances. By contrast, the wide clearances in HeartMate III induced significant disturbance to the primary flow, resulting in large incidence angle, strong secondary flows and high flow loss. At higher flow rates, the incidence angle was even larger, causing larger separation, leading to a significant decrease of efficiency and steeper performance curve compared with CH-VAD. This study shows that maglev bearings do not guarantee good blood compatibility, and more attention should be paid to the influence of secondary flows on pump performance when designing centrifugal blood pumps.<br /> (© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.)
- Subjects :
- Centrifugation
Humans
Stress, Mechanical
Blood Flow Velocity
Heart-Assist Devices
Subjects
Details
- Language :
- English
- ISSN :
- 1617-7940
- Volume :
- 23
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Biomechanics and modeling in mechanobiology
- Publication Type :
- Academic Journal
- Accession number :
- 38822142
- Full Text :
- https://doi.org/10.1007/s10237-024-01855-1