What is the MAUDE Database Telling us about 510(k) Authorization? Evaluation of Two Generations of Endovascular Arteriovenous Fistula Devices.
- Source :
-
Annals of vascular surgery [Ann Vasc Surg] 2024 Sep; Vol. 106, pp. 394-399. Date of Electronic Publication: 2024 May 28. - Publication Year :
- 2024
-
Abstract
- Background: In 2019, Bard Peripheral Vascular Inc (BV; now Becton, Dickinson and Company; Sparks, Maryland) received Food and Drug Administration (FDA) approval to begin marketing the WavelinQ EndoAVF System through a process known as 510(k) authorization. Such authorization relies on BV proving that the new WavelinQ EndoAVF System was of "substantial equivalence" to the WavelinQ 4F EndoAVF System. We set forth to analyze patient problems and device issues reported for the new device and determine if they were significantly different from the predicate device its 510(k) approval was based on.<br />Methods: FDA database Manufacturer and User Facility Device Experience was queried for all adverse report events for the WavelinQ EndoAVF System and WavelinQ 4F EndoAVF System. Data were collected on patient issues and device issues. Fisher's exact test was used.<br />Results: There were a total of 125 reports for the WavelinQ 4F EndoAVF System and 78 for the WavelinQ EndoAVF System. There was a significant increase in patient problem "hypertension" (0% vs. 5.1%; P = 0.02) for the WavelinQ EndoAVF System but a statistically significant decrease in device issue "failure to align" for the WavelinQ EndoAVF System (24.8% vs. 10.3%; P ≤ 0.01).<br />Conclusions: There were changes in device and patient outcomes between the WavelinQ EndoAVF System and WavelinQ 4F EndoAVF System. While we noted a decrease in device problem "failure to align", there was an overall increase in patients' "hypertension" rates. This highlights the importance of the FDA Manufacturer and User Facility Device Experience reporting in ensuring that device safety is maintained when devices are approved for marketing through the 510(k) process.<br /> (Published by Elsevier Inc.)
- Subjects :
- Humans
United States
Blood Vessel Prosthesis
Treatment Outcome
Blood Vessel Prosthesis Implantation instrumentation
Blood Vessel Prosthesis Implantation adverse effects
Renal Dialysis
Risk Factors
Endovascular Procedures instrumentation
Endovascular Procedures adverse effects
Device Approval
Databases, Factual
Prosthesis Design
United States Food and Drug Administration
Arteriovenous Shunt, Surgical adverse effects
Arteriovenous Shunt, Surgical instrumentation
Subjects
Details
- Language :
- English
- ISSN :
- 1615-5947
- Volume :
- 106
- Database :
- MEDLINE
- Journal :
- Annals of vascular surgery
- Publication Type :
- Academic Journal
- Accession number :
- 38815918
- Full Text :
- https://doi.org/10.1016/j.avsg.2024.03.029