Back to Search
Start Over
[Effects of Subject Position on Metal Artifact Reduction of a Reverse Shoulder Prosthesis Using Computed Tomography].
- Source :
-
Nihon Hoshasen Gijutsu Gakkai zasshi [Nihon Hoshasen Gijutsu Gakkai Zasshi] 2024 May 21. Date of Electronic Publication: 2024 May 21. - Publication Year :
- 2024
- Publisher :
- Ahead of Print
-
Abstract
- Purpose: To validate the effects of subject position on single energy metal artifact reduction (SEMAR) of a reverse shoulder prosthesis using computed tomography (CT).<br />Methods: A water phantom with a reverse shoulder prosthesis was scanned at four positions on the XY plane of the CT gantry (on-center, 50 mm, 100 mm, and 150 mm from on-center in the negative direction of the X axis, respectively). We obtained images with and without SEMAR. The artifact index (AI) was measured via physical assessment. Scheffé's (Ura) paired comparison methods were performed with the amount of metal artifact by ten radiological technologists via visual assessment.<br />Results: The AI was significantly reduced when using SEMAR. As the phantom moved away from the on-center position, the AI increased, and metal artifacts increased in Scheffé's methods.<br />Conclusion: SEMAR reduces metal artifacts of a reverse shoulder prosthesis, but metal artifacts may increase as the subject position moves away from the on-center position.
Details
- Language :
- Japanese
- ISSN :
- 1881-4883
- Database :
- MEDLINE
- Journal :
- Nihon Hoshasen Gijutsu Gakkai zasshi
- Publication Type :
- Academic Journal
- Accession number :
- 38777768
- Full Text :
- https://doi.org/10.6009/jjrt.2024-1456