Back to Search
Start Over
Severe Hemodynamic Collapse During Humerus Stabilization with Photodynamic Implant: A Report of Two Cases.
- Source :
-
JBJS case connector [JBJS Case Connect] 2024 Apr 18; Vol. 14 (2). Date of Electronic Publication: 2024 Apr 18 (Print Publication: 2024). - Publication Year :
- 2024
-
Abstract
- Case: We present 2 cases of severe hemodynamic collapse during prophylactic stabilization of impending pathologic humerus fractures using a photodynamic bone stabilization device. Both events occurred when the monomer was infused under pressure into a balloon catheter.<br />Conclusion: We suspect that an increase in intramedullary pressure during balloon expansion may cause adverse systemic effects similar to fat embolism or bone cement implantation syndrome. Appropriate communication with the anesthesia team, invasive hemodynamic monitoring, and prophylactic vent hole creation may help mitigate or manage these adverse systemic effects.<br />Competing Interests: Disclosure: The Disclosure of Potential Conflicts of Interest forms are provided with the online version of the article (http://links.lww.com/JBJSCC/C331).<br /> (Copyright © 2024 by The Journal of Bone and Joint Surgery, Incorporated.)