Back to Search Start Over

Image quality of abdominal photon-counting CT with reduced contrast media dose: Evaluation of reduced contrast media protocols during the COVID19 pandemic supply shortage.

Authors :
Layer YC
Isaak A
Mesropyan N
Kupczyk PA
Luetkens JA
Dell T
Attenberger UI
Kuetting D
Source :
Heliyon [Heliyon] 2024 Mar 18; Vol. 10 (6), pp. e28142. Date of Electronic Publication: 2024 Mar 18 (Print Publication: 2024).
Publication Year :
2024

Abstract

Rationale and Objectives: Aim of this study was to assess the impact of contrast media dose (CMD) reduction on diagnostic quality of photon-counting detector CT (PCD-CT) and energy-integrating detector CT (EID-CT).<br />Methods: CT scans of the abdominal region with differing CMD acquired in portal venous phase on a PCD-CT were included and compared to EID-CT scans. Diagnostic quality and contrast intensity were rated. Additionally, readers had to assign the scans to reduced or regular CMD. Regions-of-interest (ROIs) were placed in defined segments of portal vein, inferior vena cava, liver, spleen, kidneys, abdominal aorta and muscular tissue. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were calculated.<br />Results: Overall 158 CT scans performed on a PCD-CT and 68 examinations on an EID-CT were analyzed. Overall diagnostic quality showed no significant differences for PCD-CT with standard CMD which scored a median 5 (IQR:5-5) and PCD-CT with 70% CMD scoring 5 (4-5). (For PCD-CT, 71.69% of the examinations with reduced CMD were assigned to regular CMD by the readers, for EID-CT 9.09%. Averaged for all measurements SNR for 50% CMD was reduced by 19% in PCD-CT (EID-CT 34%) and CNR by 48% (EID-CT 56%). Virtual monoenergetic images (VMI) <subscript>50keV</subscript> for PCD-CT images acquired with 50% CMD showed an increase in SNR by 72% and CNR by 153%.<br />Conclusions: Diagnostic interpretability of PCD-CT examinations with reduction of up to 50% CMD is maintained. PCD-CT deducted scans especially with 70% CMD were often not recognized as CMD reduced scans. Compared to EID-CT less decline in SNR and CNR is observed for CMD reduced PCD-CT images. Employing VMI <subscript>50keV</subscript> for CMD-reduced PCD-CT images compensated for the effects.<br />Competing Interests: The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Yannik Christian Layer reports a relationship with 10.13039/501100004830Siemens Healthcare AG that includes: funding grants.<br /> (© 2024 The Authors.)

Details

Language :
English
ISSN :
2405-8440
Volume :
10
Issue :
6
Database :
MEDLINE
Journal :
Heliyon
Publication Type :
Academic Journal
Accession number :
38533048
Full Text :
https://doi.org/10.1016/j.heliyon.2024.e28142