- This work reports a novel POC diagnostic technique to identify the cancerous liver cell lines by designing a Source-Extended (SE) Tunnel Field Effect Transistor (TFET) having a Single-Gate (SG) with Single-Metal (SM) and Dual-Metal (DM) structure. The proposed structures have been equipped with nanocavities by trenching the gate oxide layer where the needle biopsy obtained liver sample has been immobilized. The detection is based on the difference in drain current and the ratio of the proposed device's ON and OFF state currents, which has been evaluated by obtaining the sensitivities. The cancerous and non-cancerous liver cell lines possess different dielectric properties in high frequencies ranging from 100 MHz to 5 GHz, affecting the cavity region's effective capacitances. The change in the dielectric constant of the specimen at 900 MHz has been considered which results in the change in device drain current and device performance. Various parameters of the device, like the adhesive layer in the cavity region, the material of the gate, the length of the cavities, and the orientation of the cavities, have been modified to observe the performance. The total work has been done in the simulation environment, which includes the study considering the different proportions of cancerous and non-cancerous cells in a particular specimen. A comparative analysis has been made between the performance of the proposed SM and DM gate structure. The proposed detection method has been compared with the existing methods reported in the literature. The proposed method can be considered a novel technique and can be implemented as a point of care (POC) diagnostic to detect whether the specimen liver cell line is cancerous.<br />Competing Interests: Declaration of competing interest None declared<br /> (Copyright © 2024. Published by Elsevier Ltd.)
- Subjects :
- Humans
Torso
Biopsy, Needle
Point-of-Care Systems
Liver Neoplasms diagnosis
Subjects
Details
- Language :
- English
- ISSN :
- 1873-4030
- Volume :
- 125
- Database :
- MEDLINE
- Journal :
- Medical engineering & physics
- Publication Type :
- Academic Journal
- Accession number :
- 38508806
- Full Text :
- https://doi.org/10.1016/j.medengphy.2024.104133