On the need of in vivo verifications as quality control for intraoperative electron radiotherapy in breast cancer.

Authors :
Cases C
Oses G
Herreros A
Tarrats-Rosell J
Moreno S
MollĂ  M
Source :
Clinical & translational oncology : official publication of the Federation of Spanish Oncology Societies and of the National Cancer Institute of Mexico [Clin Transl Oncol] 2024 Jul; Vol. 26 (7), pp. 1623-1629. Date of Electronic Publication: 2024 Jan 24.
Publication Year :
2024

Abstract

Introduction: Intraoperative electron radiotherapy (IOERT) is a technique aiming to deliver radiotherapy during oncological surgery. In breast IOERT, the applicator and shielding disc placement are correlated with organs at risk (OAR) irradiation, in vivo verification of these parameters is scarcely reported. The aim of our study is to report and analyze possible causes of the misalignment using radiochromic films and compare our results to others reported in the bibliography.<br />Methods: From November 2019 to April 2023, in vivo verifications were performed for 33 patients. IOERT was performed using a LIAC 10 MeV (Sordina, Italy) electron accelerator. We attached a radiochromic film to the upper side of the polytetrafluoroethylene cover of the shielding disc. The percentage of the irradiation area outside the disc was recorded and various parameters (applicator angulations, prescription depth, tumor location and breast size) were analyzed to find possible correlations.<br />Results: For 29 patients, 20 Gy were prescribed while 10 Gy were prescribed to 4 patients. The average irradiated area outside the disc was 19% (0-56%) corresponding to a surface of 4.5 cm <superscript>2</superscript> (0-17.4 cm <superscript>2</superscript> ). The applicator of 5 cm was used for most of the patients. The mean prescription depth was 1.4 cm (0.5-2.5 cm). We found no correlation between the analyzed parameters and misalignment.<br />Conclusion: This study confirms the presence and magnitude of the misalignments. We strongly recommend in vivo verifications as a quality check during IOERT procedures. The misalignment has no correlation with tumor localization parameters, so the solution could be based on technical improvements of the applicator.<br /> (© 2024. The Author(s), under exclusive licence to Federación de Sociedades Españolas de Oncología (FESEO).)

Details

Language :
English
ISSN :
1699-3055
Volume :
26
Issue :
7
Database :
MEDLINE
Journal :
Clinical & translational oncology : official publication of the Federation of Spanish Oncology Societies and of the National Cancer Institute of Mexico
Publication Type :
Academic Journal
Accession number :
38267659
Full Text :
https://doi.org/10.1007/s12094-023-03378-0