Back to Search
Start Over
Standardized in vitro bleeding tests in a non-coated novel hybrid prosthesis for frozen elephant trunk demonstrates minimal oozing during full heparinization, supported by clinical data.
- Source :
-
Frontiers in cardiovascular medicine [Front Cardiovasc Med] 2023 Dec 14; Vol. 10, pp. 1303816. Date of Electronic Publication: 2023 Dec 14 (Print Publication: 2023). - Publication Year :
- 2023
-
Abstract
- Introduction: Recent reports have questioned the blood impermeability of the novel frozen elephant trunk (FET) device E-vita Open NEO <superscript>©</superscript> (EO-NEO). Therefore, standardized in vitro bleeding tests using porcine heparinized blood were performed, as well as stress testing on the blood tightness of the collar suture line, to investigate this observation.<br />Material and Methods: EO-NEO prostheses were examined in vitro for blood permeability in three test series. Initially, antegrade perfusion with heparinized porcine blood [activated clotting time (ACT) of 500 s, with a 60 min duration] was performed, followed by ante/retrograde testing via the EO-NEO side port. Testing of the collar suture line under a tension of 10 Newton (N) within a suspension device (blood pressure 120 mmHg, ACT of 560 s, 1 min duration) was carried out with the suture material force fiber white (FFWs) yarn, using standard fixation (5 stitches/cm), FFWh yarn in hemostatic fixation (15 stitches/cm), and flow weave yarn (FWYh).<br />Results: Blood permeability testing of EO-NEO through the prosthetic lumen or via the side port demonstrated minor leakage without statistical difference between the standard and hemostatic suture lines or suture materials used, or positioning on the crimped or tapered portion ( p > 0.05). The specific collar anastomosis testing demonstrated leakage volumes of 140 ml/min for FFWs vs. 16 ml/min for FFWh ( p = 0.02), vs. 9 ml/min with the FWYh ( p = 0.01).<br />Conclusion: Different blood leakage tests showed minimal oozing and no difference in blood loss through the fabric and different collar suture lines, but unphysiological pressurized retrograde perfusion of the collar region showed significantly less leakage using FWYh and FFWh, prompting production modification of EO-NEO. Clinical results confirmed low blood loss using this novel FET device.<br />Competing Interests: HJ was a consultant for Artivion and received royalties until 12/2019. KG is a full-time employee of TEM Innovations GmbH, and DW is a full-time employee of CytoSorbents Europe. TL was employed by Medira GmbH. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest. The author(s) declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.<br /> (© 2023 Jakob, Leins, Avci-Adali, Schlensak, Wendt, Mehta, Thielmann, Görlinger, Song and Tsagakis.)
Details
- Language :
- English
- ISSN :
- 2297-055X
- Volume :
- 10
- Database :
- MEDLINE
- Journal :
- Frontiers in cardiovascular medicine
- Publication Type :
- Academic Journal
- Accession number :
- 38155987
- Full Text :
- https://doi.org/10.3389/fcvm.2023.1303816