Back to Search Start Over

On dosimetric characteristics of detectors for relative dosimetry in small fields: a multicenter experimental study.

Authors :
Casar B
Mendez I
Gershkevitsh E
Wegener S
Jaffray D
Heaton R
Pesznyak C
Stelczer G
Bulski W
Chełminski K
Smirnov G
Antipina N
Beavis AW
Harding N
Jurković S
Hwang MS
Saiful Huq M
Source :
Physics in medicine and biology [Phys Med Biol] 2024 Jan 24; Vol. 69 (3). Date of Electronic Publication: 2024 Jan 24.
Publication Year :
2024

Abstract

Objective . In this multicentric collaborative study, we aimed to verify whether the selected radiation detectors satisfy the requirements of TRS-483 Code of Practice for relative small field dosimetry in megavoltage photon beams used in radiotherapy, by investigating four dosimetric characteristics. Furthermore, we intended to analyze and complement the recommendations given in TRS-483. Approach . Short-term stability, dose linearity, dose-rate dependence, and leakage were determined for 17 models of detectors considered suitable for small field dosimetry. Altogether, 47 detectors were used in this study across ten institutions. Photon beams with 6 and 10 MV, with and without flattening filters, generated by Elekta Versa HD <superscript>TM</superscript> or Varian TrueBeam <superscript>TM</superscript> linear accelerators, were used. Main results . The tolerance level of 0.1% for stability was fulfilled by 70% of the data points. For the determination of dose linearity, two methods were considered. Results from the use of a stricter method show that the guideline of 0.1% for dose linearity is not attainable for most of the detectors used in the study. Following the second approach (squared Pearson's correlation coefficient r <superscript>2</superscript> ), it was found that 100% of the data fulfill the criteria r <superscript>2</superscript> > 0.999 (0.1% guideline for tolerance). Less than 50% of all data points satisfied the published tolerance of 0.1% for dose-rate dependence. Almost all data points (98.2%) satisfied the 0.1% criterion for leakage. Significance . For short-term stability (repeatability), it was found that the 0.1% guideline could not be met. Therefore, a less rigorous criterion of 0.25% is proposed. For dose linearity, our recommendation is to adopt a simple and clear methodology and to define an achievable tolerance based on the experimental data. For dose-rate dependence, a realistic criterion of 1% is proposed instead of the present 0.1%. Agreement was found with published guidelines for background signal (leakage).<br /> (Creative Commons Attribution license.)

Details

Language :
English
ISSN :
1361-6560
Volume :
69
Issue :
3
Database :
MEDLINE
Journal :
Physics in medicine and biology
Publication Type :
Academic Journal
Accession number :
38091616
Full Text :
https://doi.org/10.1088/1361-6560/ad154c